The stratospheric Ozone observation in Pakistan with ‘Dobson Ozone Spectrophotometer’, started in 1954, at Ozone layer monitoring station Quetta located at Geophysical center Pakistan Meteorological Department Quetta.
Dobson Ozone Spectrophotometer are the key instrument of global Ozone
observing system and form the Worlds ground based network for monitoring
atmospheric Ozone, detecting the variation of stratospheric Ozone, which
caused by natural causes and halocarbon pollutants; beside providing quality
assured Ozone measurements. These instruments are used as WMO’s standard
instrument around the world. Dobson instrument # 100 Pakistan was established
as national standard (Pakistan) at Tsukuba second international workshop
on Ozone observation and International Inter comparison of Dobson Ozone
spectrophotometer in Asia and Pacific Region (IWOAP 2) in 1996. The Pakistan
Dobson standard # 100, served as the reference instrument for home Dobson
# 43 inter comparison.
CHRONOLOGICAL RECORD OF TECHNICAL MAINTENANCE AND PROGRAME OF TOTAL OZONE OBSERVATION AND OZONE DATA-PROCESSING ON DOBSON OZONE DATA-PROCESSING ON DOBSON OZONE SPECTRPHOTOMETERS
|
|
|
|
| Original R-N tables, Q-settingTables and Vigroux 1953 Abs. Coeff. in use | 1960 | DS. GQP. AD. Obs, standard Lamp in use, lanp No. 43-1, 43-2, 43-3. |
| IAMP 1968 abs. Coeff. in use | 1968 | Dobson # 43 out of order, dt. 9-5-1964 |
| Corrected R-N tables and new Q Setting tables in use for D#100 | April 1968 | Dobson # 100 in use. |
| Corrected R-N tables and new Q-setting table in use for Dobson # 43 | March 1981 | Replacing/Removal of the HT battery upgrading system. |
| Corrected R-N tables and new Q-setting tables in use. | March 1981 | International Comparison D # 43 VS # 77 carried by Atmospheric Environment Service Toronto, Canada in Potsdam. Readjustment of the instrument. Wedge Calibration, see file "Comparisons". Standard lamp No. 43-1, 43-2, 43-3 / 100-1, 100-2, 100-3 in use. |
| June 1981 | Dobson # 100 VS # 43
Toronto / Re-calibration Upgrading # 100 VS # 43 |
|
| IAMAP 1992 abs. Coeff. in use | 1992 | Dobson # 43 in use |
| Corrected R-N tables and new Q-setting tables prepared for use. | August/ September, 1996 | International Comparison of Dobson # 100 VS Substandard Dobson # 116,
At Tsukuba, Japan.
Standard lamp No. 100 Q 1, 100 Q 2, 100 Q 3 are in use. Mercury lamp damaged in transit from Japan to Pakistan. |